Turbidity measurement (VisoTurb
Continuing with two-
point calibration
66
®
900-P)
TRB
0.2
000.0 FNU
Cal
01.07.2011
6.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
7.
Immerse the turbidity sensor in the test sample at an angle.
8.
Position the turbidity sensor in the measuring vessel.
< > < >
9.
Use
solution for each digit and confirm with
The standard solution is measured.
The measured value is checked for stability (AutoRead).
10.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
TRB
0.0
124.0 FNU
Cal
01.07.2011
11.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
12.
Immerse the turbidity sensor in the test sample at an angle.
13.
Position the turbidity sensor in the measuring vessel.
< > < >
14.
Use
solution for each digit and confirm with
The standard solution is measured.
The measured value is checked for stability (AutoRead).
15.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
1
Standard
FNU
08:00
<F2>/[
]
and
to set the concentration of the standard
2
Standard
FNU
08:00
<F2>/[
]
and
to set the concentration of the standard
®
inoLab
Multi 9310 IDS
<ENTER>
.
<ENTER>
.