®
inoLab
Multi 9310 IDS
General setting
menu for
conductivity
Setting menu
TetraCon 925
ba75937e10
08/2021
10.4 Cond measurement settings
10.4.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
tings, switch to the measured value display with
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (TetraCon 925, LR 925/01).
Default settings are printed in bold.
Menu item
Calibration /
Calibration record
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to
RS232/USB
Calibration /
Calibration interval
Reset
Menu item
Type
<ENTER>
key. After completing the set-
<M>
Possible
Explanation
setting
Displays the calibration record of
-
the last calibration.
Displays the last calibration
-
records (max. 10)
Outputs the calibration data stor-
-
age to the interface
Calibration interval for the IDS
1 ... 150 ...
conductivity sensor (in days).
999 d
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Resets all sensor settings to the
-
delivery condition (see section
10.7.1 R
MENT SETTINGS
Possible
Explanation
setting
Used measuring cell
Measuring cells whose cell con-
Cal
stants are determined through
calibration in the check- and cali-
bration standard 0.01 mol/l KCl
(1413 µS/cm @25 °C).
Calibration range:
0.450 to 0.500 cm
The currently valid cell constant
is displayed in the status line.
Settings
.
ESETTING THE MEASURE
, page 81).
-1
-
75