wtw inoLab Multi 9310 IDS Operating Manual page 37

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
Calibration evalua-
tion
ba75937e10
08/2021
Menu item
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to
RS232/USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
Display
Calibration
record
+++
++
+
-
Clean the IDS sensor according to the sensor operating manual
Error
Error
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with
 Output the displayed calibration
record to the interface with
[USB output].
 Output all calibration records to
the interface with
output].
 Quit the display with
or
 Switch directly to the measured
value display with
-
Outputs the calibration data storage
to the interface (see section 12
T
RANSMITTING DATA
Zero point
[mV]
-15 ... +15
-20 ... <-15
or
>+15 ... +20
-25 ... <-20
or
>+20 ... +25
-30 ... <-25
or
->+25 ... +30
<-30
or
>+30
<
><
>
.
<F2__>
<F1>
<ENTER>
.
<M>
.
, page 120).
Slope [mV/pH]
-60,5 ... -58,0
>-58.0 ... -57.0
-61.0 ... <-60.5
or
>-57.0 ... -56.0
-62.0 ... <-61.0
or
>-56.0 ... -50.0
<-62.0
or
> -50,0
pH value
<F2>
/
[USB
/[Back]
37