wtw inoLab Multi 9310 IDS Operating Manual page 31

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
Continuing with two-
point calibration
ba75937e10
08/2021
pH
-180.0
AutoCal TEC
AR
01.07.2011
8.
Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator) or take over the calibrated value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
9.
If necessary, finish the calibration procedure as a single-point calibra-
<M>
tion with
.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS-
pH sensor.
10.
Thoroughly rinse the sensor with deionized water.
11.
Immerse the pH sensor in buffer solution 2.
12.
When measuring without temperature sensor:
Enter the temperature of the buffer with
13.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
pH
AutoCal TEC
AR
01.07.2011
14.
Wait for the end of the measurement with stability control (beep,
[HOLD][AR] status indicator), or take over the calibrated value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
10.011
Buffer
mV
24.8
°C
08:00
<ENTER>
7.000
Buffer
0.0
mV
24.8
°C
08:00
<
><
>
.
.
pH value
31