Settings; Ph Measurement Settings; Settings For Ph Measurements - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS

Settings

Menu item
Calibration /
Calibration record
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to
RS232/USB
Calibration / Buffer
Calibration / Single-
point calibration
Calibration /
Calibration interval
Calibration / Unit for
slope
QSC / First calibration
QSC / Record of first
calibration
QSC / Control calibra-
tion
Man. temperature
ba75937e10
08/2021
10
Settings

10.1 pH measurement settings

10.1.1 Settings for pH measurements

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the
After completing the settings, switch to the measured value display with
Default settings are printed in bold.
Possible
Explanation
setting
Displays the calibration record of the last calibration.
-
Displays the last calibration records (max. 10)
-
Outputs the calibration data storage to the interface
-
Buffer sets to be used for pH calibration.
TEC
More buffers and details: see section 10.1.2 B
ConCal
NIST/DIN
FOR CALIBRATION
page 29.
...
Quick calibration with 1 buffer
yes
no
Calibration interval for the IDS pH sensor (in days).
1 ... 7 ... 999
The meter reminds you to calibrate regularly by the flashing
d
sensor symbol in the measuring screen.
Unit of the slope.
mV/pH
The % display refers to the Nernst slope of
%
-59.2 mV/pH (100 x determined slope/Nernst slope).
Starts the initial calibration with QSC buffers.
-
This menu item is only available as long as no initial calibra-
tion was carried out with the connected IDS sensor.
Displays the calibration record of the QSC initial calibration.
-
Starts the control calibration with QSC buffers.
-
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.
Entry of the manually determined temperature.
-25 ... +25 ...
This menu item is only available if an IDS adapter is con-
+130 °C
nected.
, page 70 and section 5.2
Settings
<ENTER>
key.
<M>
.
UFFER SETS
H
,
P
CALIBRATION