wtw inoLab Multi 9310 IDS Operating Manual page 61

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
Calibration evalua-
tion
Calibration record
(USB output)
ba75937e10
08/2021
Menu item
Calibration /
Calibration data
storage / Display
Calibration /
Calibration data
storage / Output to
RS232/USB
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Display
Calibration record
+++
Error
Error
Error elimination (see section 15
W
..., page 98)
HAT TO DO IF
CALIBRATION Cond
03.04.2013 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm
Sensor
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with
 Output the displayed calibration
record to the interface with
<F2>
 Output all calibration records to
the interface with
output].
 Quit the display with
[Back] or
 Switch directly to the measured
value display with
-
Outputs the calibration data storage
to the interface (see section 12
T
RANSMITTING DATA
Cell constant [cm
Within the range
0.450 ... 0.500 cm
Outside the range
0.450 ... 0.500 cm
25.0 °C
+++
Conductivity
<
><
>
.
/[USB output].
<F2__>
[USB
<F1>
/
<ENTER>
.
<M>
.
, page 120).
-1
]
-1
-1
61