Measuring The Temperature - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

ORP
Hold function
Stability control
Criteria for a stable
measured value
44
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the Stability control func-
<AR>
tion with
<ENTER>
2.
Using
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the mea-
sured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
<ENTER>
ally with
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
You can switch off the beep (see section 10.6 S
DENT SETTINGS
<ENTER>
3.
Using
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back to the pre-
vious indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
parameter
ORP
Temperature
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.

6.1.2 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera-
<AR>
<M>
or
at any time.
, activate the Stability control function manually.
at any time. If the Stability control function is
, page 79).
, start a further measurement with stability control.
Time interval
Stability in the time
interval
15 seconds
Δ : better than 0.3 mV
15 seconds
Δ : better than 0.5 °C
®
inoLab
Multi 9310 IDS
.
-
ENSOR
INDEPEN
<AR>
<M>
or
ba75937e10
-
.
08/2021

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inolab multi 9310p ids

Table of Contents