ORP

6.1

Measuring

6.1.1 Measuring the ORP

NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.
1.
Connect the ORP sensor to the meter.
The ORP measuring window is displayed.
2.
Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.
3.
Temper the test sample or measure the current temperature.
4.
Check the meter with the ORP sensor.
5.
Immerse the ORP sensor in the test sample.
U
157.0
01.02.2014
i
08:00
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 10.6.3 A
, page 80) in the System menu.
CONTROL
To start the Stability control function manually, the HOLD function must be
enabled.
mV
24.8 °C
USB output
ORP
S
UTOMATIC
TABILITY
43