Conductivity
Hold function
Stability control
Criteria for a stable
measured value
56
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 10.6.3 A
, page 80) in the System menu.
CONTROL
To start the Stability control function manually, the HOLD function must be
enabled.
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the Stability control func-
<AR>
tion with
<ENTER>
2.
Using
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the mea-
sured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
<ENTER>
ally with
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
You can switch off the beep (see section 10.6 S
DENT SETTINGS
<ENTER>
3.
Using
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time interval
parameter
Conductivity ϰ
10 seconds
Temperature
15 seconds
<AR>
<M>
or
at any time.
, activate the Stability control function manually.
at any time. If the Stability control function is
, page 79).
, start a further measurement with stability control.
Stability in the time interval
Δ ϰ: better than
1.0% of measured value
Δ : better than 0.5 °C
®
inoLab
Multi 9310 IDS
S
UTOMATIC
.
-
ENSOR
INDEPEN
<AR>
<M>
or
ba75937e10
TABILITY
-
.
08/2021