Operation
18
4.1.4 Socket field
1 IDS sensors: (pH, ORP, conductivity, D.O.)
2 USB B (device) interface
3 Power pack
4 Service interface
CAUTION
Only connect sensors to the meter that cannot return any volt-
ages or currents that are not allowed (> SELV and > current
circuit with current limiting).
WTW IDS sensors and IDS adapters meet these requirements.
4.1.5 Sensor info
You can display the current sensor data and sensor settings of a connected
sensor at any time. The sensor data are available in the measured value
<F1>
display with the
0
pH
7.007
01.02.2014
i
08:00
1.
In the measured value display:
Display the sensor data (sensor name, series number) with [
1
/[Info] softkey.
14
25.0
°C
USB output
®
inoLab
Multi 9310 IDS
4
2
3
<F1>
ba75937e10
Info].
08/2021