pH value
Stability control
Criteria for a stable
measured value
28
You can terminate the HOLD function and the Stability control func-
<AR>
tion with
<ENTER>
2.
Using
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the mea-
sured parameter stops flashing, and there is a beep.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
<ENTER>
ally with
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
You can switch off the beep (see section 10.6 S
DENT SETTINGS
<ENTER>
3.
Using
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time interval
parameter
pH value
15 seconds
Temperature
15 seconds
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.

5.1.2 Measuring the temperature

For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
<M>
or
at any time.
, activate the Stability control function manually.
at any time. If the Stability control function is
, page 79).
, start a further measurement with stability control.
Stability in the time inter-
val
Δ : better than 0.01 pH
Δ : better than 0.5 °C
®
inoLab
Multi 9310 IDS
-
-
ENSOR
INDEPEN
<AR>
<M>
or
.
ba75937e10
08/2021