Starting Dr 100S - AGFA DR 100s User Manual

Table of Contents

Advertisement

Starting DR 100s

1. Unplug the unit from the mains power.
2. Unlock the device.
• To unlock the device using the RFID reader, put your personal RFID
key on the RFID reader.
Power off button
1.
Power status indicator light
2.
RFID reader
3.
The power status indicator light starts blinking when the RFID key is
read.
• To unlock the device using the on/off keypad, press and hold the
power button until the green status indicator light starts blinking, dial
the four digit access code and press the enter button.
Numeric keypad
1.
Power on/off button
2.
Power status indicator light
3.
Keypad status indicator light
4.
Enter button
5.
If the device is successfully unlocked, the startup process continues.
• The power status indicator light lights up.
• Three beeps are heard.
• The motorized functions of the system are enabled.
• The locks to secure the DR detectors are released.
• The X-ray generator is powered on.
• The Windows log in window appears.
In case of error, the two orange leds on the battery status indicator blink
synchronously and a sound that signals error is heard.
1
2
3
1
2
3
4
5
DR 100s | Basic Workflow | 81