Adjusting the tension of the coarse focusing knob

a
b
3

Using the pre-focusing lever

a
The tension of the coarse focusing knob can be adjusted.
Insert a commercially-available flathead screwdriver into the groove c of
the tension adjustment ring. Rotating it clockwise (arrow direction) increases
the tension and counterclockwise decreases the tension of the coarse
focusing knob a .
If the stage descends by its own weight or the focus obtained
TIP
with the fine focusing knob b is lost soon, the tension is set
c
too loose. In this case, rotate the tension adjustment ring in the
arrow direction to increase the tension.
The pre-focusing lever prevents the specimen from being damaged by
collision between the specimen and objective.
1
After bringing the specimen into focus with the high magnification
objective, rotate the pre-focusing lever a in the arrow direction until
it stops.
2
To provide a certain margin for focusing range, rotate the pre-focusing
lever a by about half turn in the direction opposite to the arrow
direction from the stopped position.
If the pre-focusing lever is not positioned appropriately,
NOTE
the stage may not rise and the specimen may not be
focused. In this case, adjust the pre-focusing lever
position again.
It is recommended to use the pre-focusing lever to prevent the
TIP
specimen from damage, but if you do not use the pre-focusing
lever, rotate the pre-focusing lever in the direction opposite to
the arrow direction until it stops.