4
Optimizing Performance

When to Remove Damper and Mixer

When to Remove Damper and Mixer

The binary pump is equipped with a pressure pulsation damper and a static
mixer. The total delay volume of the pump is 600 – 800 µL (depending on system
pressure). The mixer has a volume of 400 µL.
For applications that require lowest delay volume (for example, fast gradient
methods or gradient applications with low flow rates), damper and mixer can be
bypassed.
Standard configuration
Figure 9
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Flow path modifications of the Binary Pump
Low delay volume configuration
63