Electromagnetic Field (EMF) Shielding Requirements
In order to meet current EMF compliance regulations, you must install the provided
EMI suppression ferrite on the 15-pin, shielded Generator User port cable (see also
"Generator User Port (15-Pin)" on page
1. Install the ferrite on the cable.
2. Attached it to the HFV generator.
Grounding
Do not attempt to turn on power until the generator is grounded.
The HFV generator provides an RFI ground mounting hole for a 10-32 screw. A
suitable chassis ground connection made to this mounting hole prevents or minimizes
radio frequency interference.
Connecting Cooling Water
This generator is water-cooled. Do not operate the unit until water is connected and
the cooling requirements are met.
Water connections must be made in accordance with applicable
requirements.
If you connect the cooling water on multiple generators in series, ensure
the input water temperature to all generators is less than the maximum
input water temperature.
The overtemperature fault protects only in the event of a complete loss of
coolant.
5705211-C
4-1).
Installation, Setup, and Operation
HFV 8000 Generator
5-5