Advanced Energy HFV 8000 User Manual page 113

Table of Contents

Advertisement

You can run the test only after configuring the generator in a special setup.
Note: Regular generator control is disabled when diagnostics are in progress. An
integrity report is visually indicated by the front panel LCD.
To run the diagnostics:
1. Remove the power to the generator by disengaging the circuit breaker.
2. Remove the RF output cable and cover (see
3. Replace the RF output cover in the required 180° orientation. Two microswitches
on the rear panel will sense the proper position. This provides safety from high
voltages while the test is being performed.
4. Engage the circuit breaker.
5. Ensure the RF Enable signal is disabled.
Note: You must not have RF enable set. If the RF Enable is set prior to running
diagnostics, you will receive a CANNOT RUN DIAGNOSTICS notification
on the front panel display.
6. Make sure the interlock is satisfied (that is, the front panel INTLK OK indicator
must be lit). If the indicator is off, see
7. To initiate the test, use the front panel Display Mode button to cycle through the
display modes until the display reads:
Top line—Press and Hold
Bottom line—Mode for Diag
8. Press and hold the Display Mode button for a minimum of five seconds.
9. You will receive one of two messages:
a. CANNOT RUN DIAGNOSTICS
This means the test is not set up correctly. Recheck the setup detailed above.
b. DIAGNOSTICS IN PROGRESS
10. If, after the test, the LCD displays:
a. PASS, the diagnostics routine did not find any internal problems.
b. FAIL, contact AE Global Customer Support (see
Support" on page
5705211-C
"Interlock OK Indicator Off" on page
6-10).
Installation, Setup, and Operation
HFV 8000 Generator
Figure 5-2 on page
5-4).
"AE Global Customer
6-5.
5-19