Advanced Energy HFV 8000 User Manual page 70

Table of Contents

Advertisement

®
Advanced Energy
Table 4-7. Host port commands (Continued)
Command
Description
Reads data recorded during the last diagnostic
240
test. The controller board returns 48 data
report last
diagnostic
bytes, three bytes for each of the 16 A/D
test data
channels:
• Byte 1 = Forward power test
• Bytes 2 and 3 = The forward power value
• Byte 4 = Reflected power test
• Bytes 5 and 6 = The reflected power
• Byte 7 = Inverter power test
• Bytes 8 and 9 = The inverter power value
• Byte 10 = Fast I test
• Bytes 11 and 12 = The Fast I value at the
• Byte 13 = PAED test
• Bytes 14 and 15 = The PAED value at the
4-34
4 0 = The value is in range
4 1 = The value is out of nominal
operating parameters
at the time of the test
4 0 = The value is in range
4 1 = The value is out of nominal
operating parameters
value at the time of the test
4 0 = The value is in range
4 1 = The value is out of nominal
operating parameters
at the time of the test
4 0 = The value is in range
4 1 = The value is out of nominal
operating parameters
time of the test
4 0 = The value is in range
4 1 = The value is out of nominal
operating parameters
time of the test
Communication Interfaces
Number of
Number of
Host Data
Response
Bytes
Data Bytes
0
48
5705211-C