Integration Conditions (Integ Set); Integration Mode (Mode) - YOKOGAWA WT1803 User Manual

Precision power analyzer
Table of Contents

Advertisement

Integration Conditions (Integ Set)

You can configure the following integration conditions.

Integration mode (Mode)

Integration timer (Integ Timer)
Scheduled times for real-time integration (Real-time Control)
Integration auto calibration on/off (Auto Cal)
Watt hour integration method for each polarity (WP ± Type)
Current mode for current integration (q Mode)
Rated time of integrated D/A output (D/A Output Rated Time; option)
Integration Mode (Mode)
The integration feature has the following five modes.
Integration mode
Manual integration mode
(Normal)
Normal integration mode
(Normal)
Continuous integration mode
(Continuous)
Real-time normal integration mode
(R–Normal)
Real-time continuous integration mode
(R–Continuous)
Manual integration mode
When you set the integration mode to normal integration mode (Normal) and set the integration timer to
00000:00:00, the WT1800 performs integration in manual integration mode. After integration is started, it
continues until you press Stop. However, if either of the conditions below is met, integration is stopped, and the
integration time and integrated value are held.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
Start
IM WT1801-01EN
Start
Stop
Key
Key
operation
operation
Key
Stopped by
operation
the timer
Key
Key
operation
operation
Date and
Date and
time
time
Date and
Date and
time
time
Hold
Hold
Stop
Reset
Start
9 Integrated Power (Watt hours)
Repetition
–––
–––
Repeats when
the timer expires
–––
Repeats when
the timer expires
Hold
When the maximum
integrated value is
reached
Hold
Reset
9-7