Setting Purging - YOKOGAWA ZR402G User Manual

Integrated type zirconia oxygen/humidity
Hide thumbs Also See for ZR402G:
Table of Contents

Advertisement

Default Values
n
When the analyzer is delivered or data are initialized, the parameters are by default as shown in
Table 8.20.
Table 8.20
Exhaust gas temperature
Exhaust gas pressure
8.6.5

Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
Figure 8.4
Procedure
n
Use the parameter-code table below to set the purging time.
The allowable input ranges from 0 to 60 minutes.
Table 8.21
Set item
Purging time
Default Value
n
When the analyzer is delivered, or if data are initialized, purging time is set to 0 minutes.
Parameter Codes for Exhaust Gas Temperature and Pressure Settings
Item
Default setting
101.33 kPa abs.
Displayed alternately
Display during Purging
Purging Time
Parameter code
Set range
F15
0 to 60
<8. Detailed Data Setting>
300°C
F8.5.ai
Units
minutes
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
8-20

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22aZr202gExaxt zr seriesZr202aZr22g

Table of Contents