7.5

Confirmation of Equipment Type Setting

This equipment can be used for both the Oxygen Analyzer and the Humidity Analyzer. If you
choose optional specification /HS at the time of purchase, the equipment is set for the Humidity
Analyzer.
Before setting the operating data, be sure to check that the desired model has been set.
Note that if the equipment type setting is changed after operating data are set, the operating data
that have been set are then initialized and the default settings remain. Set the equipment type
with parameter code 「F01」 . See Table 10.7 or Table 10.8, later in this manual.
CAUTION
Note that if the equipment type is changed, operation data that have already been set are
initialized (reverting to the default setting).
Table 7.3
Switch operation
>
>
>
>
>
>
>
>
The symbol [
Converter Type Setting Procedure
Display
A01
ENT
Display after the password has been entered.
F01
ENT
Touch the [∧] key to switch to Group F. If an unwanted alphabetic
character after F has been entered, continuously touch the [∧] key to
return to the original.
0
ENT
Touch the [ENT] key for confirmation. If 0 (zero) is entered, the oxygen
analyzer is already set. If 1 (one) is entered, the humidity analyzer has
been set. Change the setting following the steps below.
0
ENT
Continuously touch the [∧] key, and the position of the digit will change
from 1 to 0 to 1 to 0. Release the [ENT] key when 0 is displayed.
0
ENT
Touch the [ENT] key. The numeric value will flash.
0
ENT
Touch the [ENT] key again to stop the numeric value from flashing.
F01
ENT
Touch the [ENT] key once again, and the display will change to the
parameter code.
ENT
Basic
Touch the [>] key together with the [ENT] key to return to the basic panel
panel
display. (This is not required if you proceed to make another setting.)
display
(The displayed numeric characters indicate the measurement gas
concentration.)
] indicates that the key is being touched. Light characters indicates that the digits are flashing.