Time Constant; Setting The Device Model - GE MAC 1200 Servicing Instructions

Hide thumbs Also See for MAC 1200:
Table of Contents

Advertisement

GE Medical Systems
MAC 1200 V 1.2
Page 38
Information Technologies
227 492 21 Rev F
3.12
Time constant
From Software release V5.2 the time constant can no morenot be set in the service mode. Ist
accessable in the user mode under "System Setup" ( Allgemeine Einstellungen ), Baseline
roll filter ( Nullienienfilter ) 0.04Hz, 0.08Hz, 0.16Hz
3.13
Setting the device model
The setting of the device model is activated by pressing the "D" key, the following display
image appears:
Device
model
=
INT
{not for use in the US)
=
USA
=
Europe 2
(not for use in the US)
=
Japan
(noi for use in the US)
=
RUSS
(not for use in the US)
X=
End
Warning!
For use in the US, do not change
the device model, which
is configured
during
manufacture to device model "USA".
Changing the device model affects the selection of the output formats and languages
available, can switch over the program for Interpretation. Also affected are the default
configuration settings, e.g., override function yes/no, line filter 50/60 Hz.