Probe Recognition Check

NOTE:
After the system software loading completion, please check to ensure that the system
can recognize the probes.
Plug in the probe. In scanning mode, the probe information is displayed on the right top location of the
screen. About the probe specification for intended use on LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e,
please refer to
Plug in at least one of each type of the probes and check if each of the probes is recognized and the
probe information is displayed correctly.
8-3-6

Peripheral Device Check

Check to ensure that all the peripheral devices work properly.
For instruction of peripheral device check, please refer to
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-6 on page 3-28
Figure 8-27 Probe identification