GE
Direction 5370626-100, Revision 18
7-3-2
Check and Record the P3 Key Function
5.) In the Destinations section, record the service that is displayed.
6.) In the Physical Print Buttons section, record the parameters related to the service.
7-3-3
Setting the P3 Key to USB Quick Save
If the P3 Key is not set to USB Quick Save:
1.) While on the Connectivity screen, with the Buttons tab displayed, go to the Destinations list.
2.) From the list select USB Quick Save. Press [>>] to add the selection to the Printflow View
3.) Ensure that the Physical Print Buttons section for capture Area is set to Image Area and No
4.) The P3 Key should now be set up for USB Quick Save, sending the images directly to the USB
section.
Image Compression.
memory.
Chapter 7 Diagnostics/Troubleshooting
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
(cont'd)
7-5