Battery Performance Maintenance - GE LOGIQ e R4 Series Basic Service Manual

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
10-5-8

Battery Performance Maintenance

Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one time per year.
Please follow the flow chart below to carry out battery performance maintenance.
No
Replace battery
NOTE:
Disconnect all probes when discharge battery.
NOTE:
Discharge the battery to let the system automatically shut down.
10-10
Battery discharge
>45 minutes?
Yes
Battery charge > 3 hour
Battery power > 90%?
Yes
Battery is Ok
Figure 10-1 Flow chart of Battery Performance Maintenance
Section 10-5 - System Maintenance
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
No
Battery charge > 3 hour
Battery power > 90%?
Yes
Discharge
>45 minutes?
Yes
Battery is Ok
No
Replace battery
No
Replace battery