GE LOGIQ e R4 Series Basic Service Manual page 110

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
4-3-10-3
Doppler Mode OP Panel Controls
Table 4-10
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
4-22
Doppler Mode Touch Panel Controls
Task
Adjust Rejection
Adjust Sweep Speed
Activate Full Timeline
Select Display Format
Select Layout
Adjust Frequency
Trace Direction
Invert
Auto Calculation
Modify Calcs
Trace Method
Activate Colorize
Activate Gray Map
Dynamic Range
Adjust Angle Correct
Adjust Angle Steer
Section 4-3 - General Procedure
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Expected Result(s)
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be
processed).
Changes the speed at which timeline is
swept. Press Sweep Speed up/down to
increase/decrease the value.
Displays only timeline screen. Press the Full
Timescreen to activate.
Display layout can be preset to have B-Mode
and Time-motion side-by-side or over-under.
Display layout can be preset to have B-Mode
and Time-motion side-by-side or over-under.
Enables the adjustment of the probe's
operating frequency. Press Frequency and
select desired value. The selected frequency
is displayed in the status window.
Allows to select different trace direction.
Vertically inverts the spectral trace without
affecting the baseline position. Press invert to
invert the spectral trace. The Plus and Minus
signs on the velocity scale reverse when the
spectrum is inverted.
Enables or disables auto calculation.
Activates the window to modify the auto
calculation items.
Allows to select different trace method.
Colorize the gray scale image to enhance the
eyes' discrimination capability. Press the
Cololize, Trackball to cycle through available
maps and press Set to select.
Displays a map window adjacent to the
image. Move the trackball to select the map.
The image reflects the map as scrolled
through the selections. Press Set to select.
Controls how echo intensities are converted
to shades of gray. Click Dynamic Range to
increase/decrease the value.
Estimates the flow velocity in a direction at
an angle to the Doppler vector by computing
the angle between the Doppler vector and
the flow to be measured.
Slant the Color Flow linear image left or right
to get more information without moving
probes. Click Angle Steer to the left to slant
the linear image.
comments
For LOGIQ e, LOGIQ e
Vet and LOGIQ i
For Vivid e
For LOGIQ e, LOGIQ e
Vet and LOGIQ i