Archiving And Loading Presets - GE LOGIQ e R4 Series Basic Service Manual

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
4-3-3

Archiving and Loading Presets

NOTE:
Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an DVD-
R disk (or USB memory device) for reloading on the system.
Presets should NOT be saved on the same DVD-R disk (or USB memory device) as images. The
NOTICE
Archive Menu lists the images but does NOT list the presets stored on a DVD-R disk (or USB memory
device).
4-3-3-1
Archiving Presets to an DVD-R Disk (or USB memory device)
1.) Insert an empty (blank) DVD-R disk into the DVD-RW.
2.) Access to the Utility/Config Menu, and select System. The Backup sheet will be shown on the LCD
display.
3.) Select the item to Backup/Restore.
4.) Enter backup destination or browse through the disk to locate the destination.
5.) Select Backup now. The backup status for each item is displayed on the Result column.
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Figure 4-3 Backup Sheet
Chapter 4 Functional Checks