GE LOGIQ e R4 Series Basic Service Manual page 265

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
10-6-5-4
Data Sheet for Transducer Source Leakage Current
The test passes when all readings measure less than the values shown in
on the PM Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
CAUTION
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
.
Table 10-15 Typical Data Sheet For Transducer Source Leakage Current
Unit Power
OFF
OFF
OFF
OFF
Tester Power Polarity
Switch
ON
NORM
ON
NORM
ON
REV
ON
REV
NORM
NORM
REV
REV
Chapter 10 Care & Maintenance
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Transducer Tested:
Tester GROUND or
NUETRAL Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Table
10-11. Record all data
Measurement
10-17