4-3-8-2
M Mode OP Panel Controls
Table 4-5
Step
1
2
3
4-14
M Mode OP Panel Controls
Task
Press M Mode key
Adjust Gain
Display M-Mode Cursor
Section 4-3 - General Procedure
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Expected Result(s)
M Mode Starts
Controls the amount of echo information
displayed in an image. Turn B Mode dial to
the left/right to increase/decrease Gain. Gain
displays on the monitor in Gn (dB).
Displays the M-Mode cursor on the B-Mode
image. Press Cursor and Trackball to position
M-Mode Cursor.
Comments