Loading Presets from an DVD-R disk (or USB memory device)
1.) Insert the DVD-R disk with the archived Presets into the DVD-RW.
2.) Access to Utility->System->Backup/Restore. The Restore sheet will be shown on the LCD display.
(see
3.) Select the items needed to be stored.
4.) Select Restore. The system performs the restore and restarts.
4-3-4

Adjusting the Display Monitor

Please refer to
4-3-5

Lockout/Tagout (LOTO) requirements

Follow OSHA Lockout/Tagout requirements (USA) or local Lockout/Tagout requirements by ensuring
you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Remove battery.
6.) Control all stored and residual energy.
7.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
NOTICE
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE
GREATER THAN 30 VOLTS:
1. TURN OFF THE SCANNER.
2. UNPLUG THE SYSTEM.
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TAG
TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON
&
LOCKOUT
WILL TURN OFF.
5. REMOVE THE SYSTEM BATTERY.