Loading the System Software

NOTE:
System Software may also be referred to as the Base Image or Ghost.
While the software install procedure is designed to preserved data, you should save
WARNING
WARNING
any patient data, images, system setups and customer presets to CD, DVD, USB Flash
Drive, or USB Hard Disk before doing a software upgrade.
NOTE:
Before loading the system software, please ensure that the power can be continuously supplied
and there is no risk of power cut off during loading procedure.
There are two methods to load the system software:
Load the system software with DVD. Refer to section
DVD" on page 8-5
Load the system software with USB memory stick. Refer to Refer to section
System Software with USB Memory Stick" on page 8-10
NOTE:
When upgrading the software with DVD or USB, please remove the system from the docking
cart. Otherwise, the software update might fail.
8-4
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
from
page 8-5
to
page 8-9.
Section 8-3 - Loading the System Software
8-3-1 "Loading Base Image Software with
8-3-2 "Loading the
from
page 8-10
to
page
8-16.