Functional Checks (See Also Chapter 4) - GE LOGIQ e R4 Series Basic Service Manual

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
10-5-2

Functional Checks (See Also Chapter 4)

The functional checks take about 60 minutes to perform. Refer to the system user documentation
whenever necessary.
10-5-2-1
System Checks
Table 10-5
÷
Software Options
NOTE:
* Some software may be considered standard depending upon system model configuration.
10-6
System Functional Checks
Step
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode of
B-Mode
operation.
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that affect this
CF-Mode
mode of operation.
Verify basic Doppler operation (PW if available). Check the basic system controls that affect this mode
Doppler Modes
of operation.
M-Mode
Verify basic M-Mode operation. Check the basic system controls that affect this mode of operation.
*Applicable
Verify the basic operation of all optional modes such as Multi-Image, 3D, Harmonics, Cine,... etc.
Check the basic system controls that affect each options operation.
Xmit/Recv
Use the Visual Channel Utility on the loop connect to verify that all system xmit/recv channels are
Elements
functional.
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
LCD
Verify basic LCD display functions. Refer to Chapter 3 of the User Manual.
Software Menu
Verify Software Menu display functions. Refer to Chapter 3 of the User Manual.
check
In measurement mode, make distance measurement, get result in result window. Verify the distance
Measurements
by graduate rule. Distance Accuracy should be within ±5%. (Name result from result window Result
A, result from graduate rule Result B; Distance Accuracy= (Result B-Result A)/Result A)
Section 10-5 - System Maintenance
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Description

Advertisement

Table of Contents
loading

Table of Contents