Replacement Procedures

8-3-7 Loading the System Software

WARNING
NOTE:
NOTE:
8-3-7-1 Loading the C Disk with USB memory stick
NOTE:
NOTE:
NOTE:
8-8
While the software install procedure is designed to
preserved data, you should save any patient data,
images, system setups and customer presets to CD,
DVD, USB Flash Drive, or USB Hard Disk before doing a
software upgrade.
Before loading the system software, please ensure that the
power can be continuously supplied and there is no risk of
power cut off during loading procedure.
Before loading/reload the system software, please change
system language to English. Otherwise, there will be garbage on
the system after software loading/reload.
There is one method to load the system software:
Load the system software with USB memory stick.
While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducer.
Please ensure AC adapter is connected during system upgrade!
1. Insert the USB memory stick labeled "System & Application
Software" to the system.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB
memory stick automatically.
LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8