Loading the System Software

WARNING
NOTE:
WARNING
Loading the System Software with USB memory stick/DVD
NOTE:
NOTE:
NOTE:
NOTE:
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
While the software install procedure is designed to
preserved data, you should save any patient data,
images, system setups and customer presets to CD,
DVD, USB Flash Drive, or USB Hard Disk before doing a
software upgrade.
Before loading the system software, please ensure that the
power can be continuously supplied and there is no risk of
power cut off during loading procedure.
There are two methods to load the system software:
Load the system software with USB memory stick.
Load the system software with DVD.
Before software upgrade, be sure to DISCONNECT all the
peripherals from the system, otherwise the software upgrade
will fail or the system cannot work properly after software
upgrade.
Before starting this procedure, remove all probes and
peripherals and remove them from the Docking Cart.
While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducer.
Please ensure AC adapter is connected during system upgrade!
1. Insert the Upgrade DVD labeled "System & Application
Software" in the DVD-RW or insert the USB memory stick to
the USB port of the system.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB
memory stick/DVD automatically.