USING THE SYSTEM
5.2.2

Turning On System

How to Turn On
Steps
Waiting Period
System User Guide Rev. 1.0 Curve Dual Display, Curve Single Display Ver. 1.0
1.
Ensure that all required cables have been connected at the connection panel (see
page 49).
2.
Plug in mains power at Monitor Cart (page 59).
3.
If using wireless camera communication with Curve Dual Display: Plug in mains power
at Camera Cart (see page 70).
4.
Turn on mains power switch
5.
Wait until the blue mains power LED is lit and solid.
6.
Press on/off button
a
If system has been turned off, wait at least 60 seconds before turning it on again. Oth-
erwise, high input current may trip the circuit breaker.
.
f
to start the system.