Panels and System Ventilation
3.4.4
LED Indications
Mains Power LED
Color
Blue
Power On/Off LED
Color
Green
UPS LED
Color
Orange
Error LED
Color
Red
CD/DVD LED
Color
Green
56
Status
Explanation
Off
Mains power is turned off. Safe to unplug system.
Mains power is on. Electronic components are off (press power
Solid
button to turn them on).
Status
Explanation
Off
Electronic components are off.
Electronic components are booting. Wait until LED is solid before
Flashing
use.
Solid
Electronic components are on. System is ready for use.
Status
Explanation
Off
System is not running on UPS.
System is running on UPS.
Flashing
If battery is fully charged, system can run for five minutes on the
UPS. After that time it will shut down automatically.
Solid
Battery is being charged.
Status
Explanation
Off
No malfunctions detected.
Malfunction has been detected (see page 33). Contact Brainlab
Solid
support.
Status
Explanation
Off
CD/DVD not being accessed or no CD/DVD inserted.
Flashing
CD/DVD is being accessed.