Advertisement

Parking and Storage
6.5.3

Long Term Storage

General
Before and after placing the system into storage for an extended period of time, steps must be taken
Information
to ensure proper functioning.
Storage for 1-6
Before storing system for one to six months:
Months
Steps
Before using equipment that has been stored for approximately three months:
Steps
Storage for
Before storing system for six months or longer:
6 Months or Longer
Steps
Before using equipment that has been stored for six months or longer:
Steps
Disposal
If you do not intend to use the system again, please follow the disposal information on page 9.
118
1.
Make sure that the UPS battery is fully charged.
2.
Unplug all external cabling and store it together with the equipment.
3.
Cover camera and monitors with protection covers.
4.
Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
1.
Clean the equipment (see page 120).
2.
Fully recharge UPS battery (see page 31).
3.
Start system and verify correct function before using it for surgery (see page 88).
1.
Contact Brainlab support to remove the UPS Battery from the equipment.
2.
Unplug all external cabling and store it together with the equipment.
3.
Cover camera and monitors with protection covers.
4.
Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
1.
Clean the equipment (see page 120).
2.
Contact Brainlab support to install a new UPS battery.
3.
Fully recharge UPS battery (see page 31).
4.
Perform a recurrent test to verify safety and effectiveness of equipment (offered by
Brainlab, see Technical User Guide).
5.
Start system and verify correct function before using it for surgery (see page 88).