CAMERA CART
4.4

Camera Use

4.4.1

Overview

General
The Camera Carts:
Information
• Enable mobile positioning of camera
• Allow adjustment of angle and height of camera
• Can be attached via cable to the Monitor Cart (or communicate wirelessly with Curve Dual
Safe Use
Only one person should operate the Camera Cart at any given time.
Before use, ensure that the camera and all parts of the Camera Cart are functioning properly and
there is sufficient space for camera movement.
Compatibility
Only use the Camera Cart with approved Brainlab navigation systems.
Only use the Sterile Camera Handle Sleeve specified by Brainlab for the Camera Cart.
Absolute
The camera is not intended for absolute measurements. The camera is designed to provide the rel-
Measurements
ative location of trackable instruments.
System User Guide Rev. 1.0 Curve Dual Display, Curve Single Display Ver. 1.0
Display)
To avoid risk of electrical shock, do not touch patient, Camera Cart, and/or camera si-
multaneously.
Do not step or sit on Camera Cart base.
Do not attach or hang anything on the Camera Cart, due to risk of the Camera Cart tip-
ping over.
Do not use the Camera Cart if cables are visibly damaged.
When adjusting the Camera Cart, take great care not to tilt it to a degree where there
is a chance it could tip over.
Do not place Camera Cart on uneven surfaces. Otherwise, it could move on its own,
resulting in serious damage.
Do not make any changes to the camera components. Changes to the equipment may
cause severe harm to the patient, user or 3rd party.
71