CAMERA CART
Safe Handling
How to Proceed
without Motorized
Adjustment
If motorized camera movement fails (e.g., non-responsive or uncontrolled movement) during a pro-
cedure, turn off camera on/off switch
You can continue the surgery, using manual camera adjustment if necessary (see page 77). Track-
ing is not affected.
System User Guide Rev. 1.0 Curve Dual Display, Curve Single Display Ver. 1.0
Ensure that sufficient space has been given to allow for motorized camera movement
during the procedure. Supervision is required during camera movement to ensure the
camera will not come into contact with equipment or people.
Verify that motorized camera adjustment is functioning properly before beginning sur-
gery.
Do not use motorized camera adjustment if cables are visibly damaged.
Do not position camera over patient.
When adjusting the camera position be aware that the Camera Cart or parts thereof
may move unintentionally.
The components in the camera unit are precise, sensitive optical instruments. Drop-
ping or jolting the camera may cause miscalibration.
Do not touch the camera lenses. Smudges on lenses can distort tracking.
a
Figure 33
a
and contact Brainlab support.
a
81