SYSTEM OVERVIEW
2.3

System Setup

2.3.1

Overview

General
The Curve system is an optical-tracking, touchscreen-operated planning and navigation system de-
Information
signed for pre- and intraoperative use.
• Monitor Cart and Camera Cart are delivered pre-installed and ready for use.
• All system components are suitable for continuous use during surgical procedures.
• Only use system components indoors, in an operating room.
How to Safely
Position the
Steps
System
Sterile Field
The Curve system is unsterile.
Setup for Wireless
In the ideal setup for wireless camera communication, the two carts have a direct line of sight to each
Communication
other. If people are positioned directly between the antennas, wireless connection may be negatively
influenced.
If wireless camera communication appears to be unstable, use cable-based communication.
System User Guide Rev. 1.0 Curve Dual Display, Curve Single Display Ver. 1.0
1.
Remove protection covers before entering the OR.
2.
Move equipment into the OR
3.
Angle the system so that:
• The Curve logo on the Monitor Cart points toward the patient
• User and connection panels and the Camera Cart connector point away
from the patient
NOTE: Do not position the camera, monitors or any other part of the Curve
system directly over the patient.
4.
Lock all brakes on the Monitor Cart and Camera Cart.
5.
Move monitors into desired position.
6.
Lock all locks on the monitor arms.
7.
Move camera into desired position.
8.
Connect all cables at the connection panel.
Ensure that the system is set up so it is not possible for the patient to touch or come
in contact with the equipment.
As the Camera Carts do not offer a mains power switch, ensure that the system is set
up in a way that the mains power plug (on the Monitor Cart) is easily accessible. In
case of Camera Cart malfunction, you must be able to easily unplug the mains power
cable.
Do not allow the camera, monitors or any other part of the Curve system to enter the
sterile field.
See
Page 101
Page 114
Page 23
Page 98
Page 44
Page 46
Page 77
Page 50
21