Operating Position - Brainlab CURVE DUAL System User's Manual

Hide thumbs Also See for CURVE DUAL:
Table of Contents

Advertisement

Camera Use
4.4.3

Operating Position

How to Put Camera
Cart into Operating
Steps
Position
Positioning
In case of interference with other devices caused by the IR light emitted by the camera, reposition
Considerations
the camera and/or affected device until there is no longer interference.
Safe Positioning
76
1.
Ensure that Camera Cart is in transport position (see page 107).
2.
Using the transport handle, move Camera Cart to desired location.
3.
Lock all four wheel brakes (see page 100).
4.
Remove camera protection cover.
5.
Grip camera handle and move camera arm to desired position
NOTE: The gas spring of the camera joint keeps the camera in desired position.
6.
If applicable, plug camera cable into Monitor Cart.
7.
Attach Sterile Camera Handle Sleeve to camera handle.
Reflections of IR light, e.g., from drapes or shiny surfaces, may interfere with the cam-
era's ability to correctly track instruments. The same may be true of the camera's IR
light, which can interfere with other devices, e.g., pulse oximeter, remote control sys-
tems or IR sensitive microscopes.
Lock all four wheel brakes when Camera Cart is in operating position.
Do not place Camera Cart or any of its parts over the patient.
System User Guide Rev. 1.0 Curve Dual Display, Curve Single Display Ver. 1.0