Figure 2-8 Sensor (Average Path) Functional Test Setup With Attenuator - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

Figure 2-8 Sensor (Average Path) Functional Test Setup With Attenuator

7 Turn on the power reference:
Press
8 Switch to the power sensor's lower range:
• Press
• Scroll through the Range values by pressing
LOWER.
• A pop- up window is displayed.
• Press
9 Verify that the display reads − 30 dBm ±1 dB. Record the reading in
Table 2-
10 Switch to the power sensor's upper range:
• Press
• Move to the Range parameter using the up/down arrow keys.
• Scroll through the Range values by pressing
UPPER.
• Press
11 Verify that the display reads − 30 dBm ± 1 dB. Record the reading in
Table 2-
12 Turn on the power reference:
Press
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power Meter
Power Sensor
30 dB attenuator
CHANNEL A
Power sensor cable
Zero
,
Power Ref On
Cal
.
Channel
.
Done
6.
.
Channel
.
Done
6.
Zero
,
Power Ref Off
Cal
Performance Tests
Change
. Select a value of
Change
. Select a value of
2
45