Agilent Technologies E4416A Service Manual page 177

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

D
Measurement Uncertainty Analysis - Instrument Accuracy Test
Worst case error on 15.8 V maximum power setting
Error = ((10 ppm × 15.8 V) + (0.3 ppm × 100.0 V)) + (0.5 ppm × 15.8 V) +
5((0.15 ppm × 15.8 V) + (0.1 ppm × 100 V))
= 257.75 μV
Percentage Error = 0.00163%
Voltage Error
257.75
The relationship between the applied voltage from the Universal Source to the
corresponding power is given from:
1 mW = 0.14493 mV
This is a linear relationship, for example, 0.1 mW is equal to 0.014493 volts.
This relationship is valid between 0 dBm and
However, above 0 dBm a look up table is used for voltage to mW conversion.
This is due to the sensors non linear behavior at higher power levels. The
voltage correction at the appropriate points is shown in
Table D-2 Voltage Correction Multiplier
Power (mW)
1
10
100
D-156
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Distribution
μ
V
Rectangular
Standard Uncertainty
μ
148.8
V
30 dBm.
Table D-
Voltage Correction Multiplier
1
1.006577
1.075533
E4416A/E4417A Power Meters Service Guide