Post-Repair Adjustments

Table 3- 1
assemblies.

Table 3-1 Post Repair Adjustments, Tests, and Checks

Assembly Replaced
A1 Power Supply
A2 Processor Assembly
A3 Front Panel Assembly
A4 Comms Assembly
A5 Daughter Assembly
A6 Measurement Assembly
W1 or W2 Power Sensor Cable
BT1 Lithium Battery
Power Reference Cable Kits
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
lists the adjustments related to repairs or replacement of any of the
Adjustments
Related Adjustments, Performance
Tests or Self Tests
Self Test
Power Reference Frequency and Level
Test/Adjustment. Display Brightness and
Contrast Adjustment.
Instrument Accuracy Test, Display
Brightness and Contrast Adjustment.
All automated Tests and Serial Interface
Self Test.
Instrument Accuracy Test.
Instrument Accuracy Test.
Instrument Accuracy Test.
Battery Self Test.
Power Reference Level Test.