Agilent Technologies E4416A Service Manual page 175

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

D
Measurement Uncertainty Analysis - Instrument Accuracy Test
3458A DMM Uncertainty:
The uncertainty of the voltage measured by the 3458A is:
Error = (ppm of reading + ppm of range) + (factory traceability) + (temperature
coefficient)
Conditions: 1 Year Calibration and temperature variance +/
dc volts.
Reference: 3458A Operating Manual Appendix A
Factory Traceability: From 6 ppm to 0.5 ppm Standard Uncertainty depending on
range.
Worst case error on 0.1 V range:
Error = ((9 ppm × 100 mV) + (3 ppm × 100 mV)) + (6 ppm × 100 mV) + 5((0.15
ppm × 100 mV) + 100 mV))
= 2.375 μV
Percentage Error = 0.002375%
Voltage Error
μ
2.375
D-154
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Distribution
V
Rectangular
5 °C with ACAL on
Standard Uncertainty
μ
1.371
V
E4416A/E4417A Power Meters Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4417a

Table of Contents