Procedure

1 Ensure that the power meter has been powered on for at least 30 minutes
before making any measurements.
2 Connect the equipment as shown in
measure frequency.
3 Turn the power reference on by pressing:
4 Observe the reading on the frequency counter.
If it is 50.0 ±0.5 MHz, no adjustment of the power reference oscillator
frequency is necessary. If it is not within these limits, adjust the power
reference oscillator frequency as described in step 5 and step 6.
5 Remove the power meter cover.
6 Adjust A2L9 to obtain a 50.0 ±0.5 MHz indication on the frequency counter.
Refer to
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Zero
,
Power Ref On
on the Agilent E4416A.
Cal
Zero
More
,
Cal
,
,
Power Ref On
Cal
Figure 3- 2
for the position of A2L9.
Adjustments
Figure 3-
1. Set up the counter to
on the Agilent E4417A.
3
63