Repeatability; Table C-1 Measurement Repeatability Results - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

C
Measurement Uncertainty Analysis - Power Reference Level Test
Therefore,
U R ( )
(
=
0.0036
0.0046 Ω
=

Repeatability

To verify the overall repeatability of the measurement, the test was repeated
20 times, using the same DUT. For each test cycle, the Power Reference Level
Test and the Power Reference Level Adjust tests, were carried out; at no point
during these tests was the 1mW on the DUT adjusted. The thermistor was
disconnected and reconnected between each test with the sensor position
rotated each time. The results for this evaluation are shown in

Table C-1 Measurement Repeatability Results

Measurement
1
2
3
4
*
5
6
7
8
9
10
11
12
C-144
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
×
(
)
200
10
+ +
5
3
1000
=
---------------------------------------------
+
-------------------------------------------- -
1000000
)
+
0.0010
ADJUST REF.
0.999548951
0.999541333
0.999449078
0.999546994
0.999138526
0.999647055
0.999539391
0.999567191
0.999600963
0.999405584
0.999523392
0.999504386
×
(
)
0.5
+
0.5
1000000
REF. LEVEL TEST
0.999499495
0.999554257
0.999414037
0.999509401
0.9991726
0.999454911
0.999515949
0.999558795
0.999554633
0.999377721
0.999566084
0.999504323
E4416A/E4417A Power Meters Service Guide
Table C-
1.