Table 2-5 Power Reference Level Test Result - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

2
Performance Tests
11 Round the DVM reading to two decimal places.
Record this reading as V
• V
12 Disconnect the DVM negative input lead from the VRF connector on the
Agilent 432A. Reconnect it to the Agilent 432A chassis ground.
13 Observe the DVM reading. Record the reading as V
• V
14 Calculate the Power Reference Oscillator power using equation 2- 4.
15 Verify that P
reading in the table.

Table 2-5 Power Reference Level Test Result

40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
.
1
___________
1
___________
comp
is within the limits shown in
meas
Min
0.991 mW
.
comp
Table 2- 5
and record the
Max
Actual Result
1.009 mW
__________________
E4416A/E4417A Power Meters Service Guide