2
Performance Tests
11 Round the DVM reading to two decimal places.
Record this reading as V
• V
12 Disconnect the DVM negative input lead from the VRF connector on the
Agilent 432A. Reconnect it to the Agilent 432A chassis ground.
13 Observe the DVM reading. Record the reading as V
• V
14 Calculate the Power Reference Oscillator power using equation 2- 4.
15 Verify that P
reading in the table.
Table 2-5 Power Reference Level Test Result
40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
.
1
___________
1
___________
comp
is within the limits shown in
meas
Min
0.991 mW
.
comp
Table 2- 5
and record the
Max
Actual Result
1.009 mW
__________________