Combined Standard Uncertainty uc(y):

There is no correlation associated with this measurement
Hence:
2
y ( )
uc
Table D- 3
from 3 μW (
Table D-3
Combined Standard Uncertainty
Power
μ
3.16
W
0.0097
μ
μ
10
W
0.009
μ
31.6
W
0.0095
μ
μ
100
W
0.035
μ
μ
316
W
0.03
1 mW
0.00007 mW
3.16 mW
0.00022 mW
10 mW
0.00069 mW
31.6 mW
0.0022 mW
100 mW
0.0092 mW
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Measurement Uncertainty Analysis - Instrument Accuracy Test
2
[
(
)
]
,
=
C
u TSE
TSE
shows the combined standard uncertainty at all the power levels
25 dBm) to 100 mW (20 dBm).
U(TSE)
U(MV) (Converted)
μ
μ
W
0.00946
W
μ
W
0.00946
W
μ
μ
W
0.00946
W
μ
W
0.00946
W
μ
W
0.042
W
0.000042 mW
0.000389 mW
0.000389 mW
0.00095 mW
0.00095 mW
2
[
(
)
]
C
u MV
MV
uc(y)
μ
0.0135
W
μ
0.0131
W
μ
0.0134
W
μ
0.0363
W
μ
0.0516
W
0.000082 mW
0.000447 mW
0.000792 mW
0.00239 mW
0.00924 mW
D
k uc(y)
k = 2
μ
0.0271
W
μ
0.0262
W
μ
0.0268
W
μ
0.0725
W
μ
0.1032
W
0.00016 mW
0.00089 mW
0.00158 mW
0.00479 mW
0.01850 mW
D-159