5
Replaceable Parts
After replacing a processor board, the display brightness and contrast must be
N O TE
adjusted. For further information refer to
Adjustment".
Firmware should be downloaded to the instrument after the processor board is
N O TE
replaced. Refer to
98
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter
"Firmware Upgrades"
on page 94.