Table 2-4 Instrument Accuracy Results - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

13 Calibrate the power meter:
Press
14 Set the DC calibrator output to the values shown in
setting, set the filter length of the power meter to the carbon- dating value,
and verify that the power meter reading is within the limits shown. Refer to
step 10 for details on how to setup the filter length.

Table 2-4 Instrument Accuracy Results

DC
Effective
Power
Calibrator
Power
Meter
Setting
Applied
Filter
Setting
458.31 μV
3.1623 μW
256
10 μW
1.4493 mV
256
31.6228 μW
4.5831 mV
256
14.493 mV
100μW
64
316.2278 μW
45.831 mV
64
144.93 mV
1 mW
16
459.12 mV
3.1623 mW
16
1.4588 V
10 mW
16
4.6872 V
31.6228 mW
16
15.588 V
100 mW
16
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Zero
Cal
,
.
Cal
Meas Unc Lower Limit Ch A
Result
0.028 μW
3.0977 μW
_____
0.027 μW
9.904 μW
_____
0.027 μW
31.428 μW
_____
0.073 μW
99.49 μW
_____
0.104 μW
314.72 μW
_____
0.2 μW
0.9954 mW
_____
0.9 μW
3.1477 mW
_____
1.6 μW
9.954 mW
_____
4.8 μW
31.477 mW
_____
18.5 μW
99.54 mW
_____
Performance Tests
Table 2-
4. For each
Ch B
Upper Limit GuardBa
Result
nd Lower
Limit (in
μW)
3.2269 μW
_____
3.1257
10.096 μW
_____
9.931
31.819 μW
_____
31.455
100.51 μW
_____
99.563
317.74 μW
_____
314.824
_____
1.0047 mW
995.6
_____
3.1769 mW
3148.6
_____
10.046 mW
995.6
_____
31.769 mW
31481.8
_____
100.46 mW
99558.5
2
GuardBa
nd Upper
Limit (in
μW)
3.1989
10.069
31.792
100.437
317.636
1004.5
3176
10044.4
31764.2
100441.5
35